Z-2149-2011
4Z1c Matrix Array Transducer, model number 10033682, used with ACUSON SC2000 ultrasound systems, model number...
Siemens Medical Solutions USA, Inc. / initiated 2009-07-10 / Terminated / root cause: manufacturing process
Siemens Medical Solutions USA, Inc. began a recall of 4Z1c Matrix Array Transducer, model number 10033682, used with ACUSON SC2000 ultrasound systems, model number 10433816, manufactured by Siemens Medical... on . The reason FDA records is "Burn risk-- due to coolant leaks, patients may suffer burns or blisters due to increased temperature". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2149-2011
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- 4Z1c Matrix Array Transducer, model number 10033682, used with ACUSON SC2000 ultrasound systems, model number 10433816, manufactured by Siemens Medical Solutions, Ultrasound Business Unit, Mountain View, CA. The 4Z1c transducer is a special purpose ultrasound transducer designed to help clinicians perform adult echocardiography and pre-operative and intra-operative surgical assessments. Secondary applications include pediatric echocardiography, primarily benefiting older and larger pediatric populations, general abdominal and intra-operative pediatric abdominal imaging and peripheral vessel imaging.
- Reason for recall
- Burn risk-- due to coolant leaks, patients may suffer burns or blisters due to increased temperature.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- A customer safety advisory notification was sent out to customers on 7/10/09, and a mandatory software update will also be conducted.
- Quantity in commerce
- 42 units
- Distribution
- Nationwide. Foreign: Japan, Italy, South Korea, Singapore, Yemen, Germany, Australia, and Greece.
- Product codes and lots
- seial numbers
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.