Z-1012-2010

PTS Panels CHOL+HDL+GLU Panel test strips for professional use with CardioChek P-A analyzers, Manufactured by...

Polymer Technology Systems / initiated 2009-08-04 / Terminated

Polymer Technology Systems, Inc. began a recall of PTS Panels CHOL+HDL+GLU Panel test strips for professional use with CardioChek P-A analyzers, Manufactured by Polymer Technology Systems, Inc., Indianapolis, IN on . The reason FDA records is "Possible under recovery of cholesterol, HDL and glucose. A loss of recoverable cholesterol and HDL cholesterol is a manageable to moderate health risk. The clinical applications analysis shows that lot 1891 shows such a...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1012-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Polymer Technology Systems
Product
PTS Panels CHOL+HDL+GLU Panel test strips for professional use with CardioChek P-A analyzers, Manufactured by Polymer Technology Systems, Inc., Indianapolis, IN.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
Possible under recovery of cholesterol, HDL and glucose. A loss of recoverable cholesterol and HDL cholesterol is a manageable to moderate health risk. The clinical applications analysis shows that lot 1891 shows such a loss of activity. A loss of recoverable Glucose measurement is a moderate to high risk, based upon the degree of under-recovery, and the frequency of occurrence.
FDA root cause
Other
Action
Consignees were notified by letter dated 7/28/09 to discontinue use of, and to destroy, the reagent and to contact the firm for replacement product. For questions, please contact the company Customer Service toll-free at 877-870-5610 or + 1-317-870-5610 (outside the US).
Quantity in commerce
962
Distribution
Nationwide.
Product codes and lots
Lot I891.