Z-0143-2010

Power cord for Optional Thermal Printer (For use with Model 3300 COC), Oximetrix 3 SO2/CO Computer, Q-Vue CCO...

Abbott Medical / initiated 2009-08-11 / Terminated / root cause: design

Hospira Inc began a recall of Power cord for Optional Thermal Printer (For use with Model 3300 COC), Oximetrix 3 SO2/CO Computer, Q-Vue CCO Computer and Q2 CCO/SO2 Computer, List Number... on . The reason FDA records is "Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock hazard". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0143-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Power cord for Optional Thermal Printer (For use with Model 3300 COC), Oximetrix 3 SO2/CO Computer, Q-Vue CCO Computer and Q2 CCO/SO2 Computer, List Number 50132-04-05, AC cord manufactured by Electri-Cord, Westfield PA, Infusion pumps manufactured by Hospira, Morgan Hill, CA.
Product code
dxg
Reason for recall
Fire/Shock hazard-- The power cord used in the device may cause sparks/flashes and poses a fire and shock hazard.
Root cause tag
design
FDA root cause
Component design/selection
Action
Hospira initiated its recall on 08/14/2009. A nationwide press release was issued to AP, and a letter to consignees was sent via UPS. The firm will replace affected cords and effectiveness checks will be conducted through tracking of servicing.
Quantity in commerce
204,999 power cords distributed for all devices
Distribution
Worldwide distribution.
Product codes and lots
AC power cords, all units of Part Numbers: 826-85559-001,  826-85559-003,  826-85559-008,  826-85559-010,  826-96057-003