Z-0127-2010

This device is not approved in the United States. Medtronic Concerto, C174AWK. Dual Chamber implantable...

Medtronic, Inc. / initiated 2009-09-09 / Terminated / root cause: manufacturing process

Medtronic Inc. Cardiac Rhythm Disease Managment began a recall of This device is not approved in the United States. Medtronic Concerto, C174AWK. Dual Chamber implantable cardioverter defibrillator with cardiac... on . The reason FDA records is "A subset of active implanted Concerto CRT-D and Virtuoso ICD devices worldwide (estimated at 6,300) may not meet expected device longevity due to gradually increasing current drain caused by low voltage capacitor...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0127-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
This device is not approved in the United States. Medtronic Concerto, C174AWK. Dual Chamber implantable cardioverter defibrillator with cardiac resynchronization therapy, atrial and ventricular therapies. Medtronic, Inc., Minneapolis, MN 55432 USA.
Product code
NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
Reason for recall
A subset of active implanted Concerto CRT-D and Virtuoso ICD devices worldwide (estimated at 6,300) may not meet expected device longevity due to gradually increasing current drain caused by low voltage capacitor degradation. This issue may present in the affected devices as reaching the Recommended Replacement Time (RRT) earlier than projected. This issue does not compromise device functionalit
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
An "Important Medical Device Information" letter, dated September 2009 was hand delivered by Medtronic Representatives beginning 09/09/09. The letter describes the issue, identifies affected devices and provides recommendations for patients with devices in the affected subset.. A Patient letter was mailed to patients beginning 09/28/09. For additional information: Doctors are advised to contact your local Medtronic Representative or Medtronic Technical Services at 800-723-4636; Device tracking managers are advised to contact your Medtronic representative, or the Patient Registration Department at 1-800-328-2518 x41832; and Patients are advised to contact Medtronic Patient Services number at 800-551-5544 ext. 41835.
Quantity in commerce
267
Distribution
Worldwide distribution (except Idaho and Wyoming) including countries of Australia, Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Norway, Portugal, Saudi Arabi, Spain, Sweden, Switzerland, Turkey and United Kingdom.
Product codes and lots
serial numbers: PVU600353S, PVU600444S, PVU600455S, PVU600466S, PVU600476S, PVU600478S, PVU600494S, PVU600497S, PVU600502S, PVU600505S, PVU600506S, PVU600507S, PVU600508S, PVU600512S, PVU600516S, PVU600518S, PVU600520S, PVU600522S, PVU600526S, PVU600529S, PVU600531S, PVU600534S, PVU600535S, PVU600536S, PVU600539S, PVU600543S, PVU600549S, PVU600558S, PVU600561S, PVU600568S, PVU600571S, PVU600573S, PVU600574S, PVU600582S, PVU600585S, PVU600595S, PVU600597S, PVU600598S, PVU600605S, PVU600607S, PVU600612S, PVU600617S, PVU600618S, PVU600619S, PVU600621S, PVU600622S, PVU600627S, PVU600631S, PVU600638S, PVU600640S, PVU600641S, PVU600646S, PVU600648S, PVU600652S, PVU600660S, PVU600664S, PVU600665S, PVU600675S, PVU600679S, PVU600680S, PVU600681S, PVU600682S, PVU600683S, PVU600685S, PVU600692S, PVU600694S, PVU600695S, PVU600698S, PVU600699S, PVU600703S, PVU600706S, PVU600708S, PVU600709S, PVU600711S, PVU600713S, PVU600718S, PVU600719S, PVU600721S, PVU600722S, PVU600723S, PVU600729S, PVU600767S, PVU600768S, PVU600781S, PVU600800S, PVU600806S, PVU600807S, PVU600808S, PVU600816S, PVU600819S, PVU600823S, PVU600825S, PVU600833S, PVU600834S, PVU600837S, PVU600840S, PVU600892S, PVU600898S, PVU600917S, PVU600949S, PVU600951S, PVU600957S, PVU600961S, PVU601001S, PVU601011S, PVU601012S, PVU601015S, PVU601020S, PVU601034S, PVU601035S, PVU601044S, PVU601052S, PVU601054S, PVU601058S, PVU601085S, PVU601108S, PVU601110S, PVU601115S, PVU601117S, PVU601120S, PVU601124S, PVU601133S, PVU601136S, PVU601140S, PVU601141S, PVU601142S, PVU601146S, PVU601147S, PVU601151S, PVU601153S, PVU601154S, PVU601159S, PVU601160S, PVU601162S, PVU601163S, PVU601164S, PVU601168S, PVU601170S, PVU601171S, PVU601177S, PVU601183S, PVU601190S, PVU601195S, PVU601197S, PVU601210S, PVU601211S, PVU601214S, PVU601219S, PVU601228S, PVU601231S, PVU601233S, PVU601234S, PVU601241S, PVU601243S, PVU601249S, PVU601256S, PVU601257S, PVU601258S, PVU601263S, PVU601269S, PVU601270S, PVU601274S, PVU601276S, PVU601282S, PVU601284S, PVU601285S, PVU601287S, PVU601289S, PVU601293S, PVU601301S, PVU601304S, PVU601314S, PVU601317S, PVU601320S, PVU601322S, PVU601325S, PVU601326S, PVU601329S, PVU601333S, PVU601341S, PVU601342S, PVU601343S, PVU601345S, PVU601350S, PVU601351S, PVU601354S, PVU601366S, PVU601368S, PVU601381S, PVU601398S, PVU601405S, PVU601410S, PVU601413S, PVU601430S, PVU601432S, PVU601433S, PVU601437S, PVU601456S, PVU601495S, PVU601499S, PVU601505S, PVU601519S, PVU601520S, PVU601521S, PVU601523S, PVU601524S, PVU601528S, PVU601532S, PVU601533S, PVU601548S, PVU601551S, PVU601555S, PVU601573S, PVU601577S, PVU601578S, PVU601581S, PVU601582S, PVU601585S, PVU601586S, PVU601592S, PVU601593S, PVU601604S, PVU601607S, PVU601611S, PVU601612S, PVU601614S, PVU601620S, PVU601623S, PVU601625S, PVU601635S, PVU601642S, PVU601651S, PVU601677S, PVU601686S, PVU601688S, PVU601694S, PVU601718S, PVU601719S, PVU601721S, PVU601723S, PVU601836S, PVU613075S, PVU613083S, PVU613107S, PVU613111S, PVU613112S, PVU613119S, PVU613123S, PVU613125S, PVU613127S, PVU613128S, PVU613129S, PVU613139S, PVU613141S, PVU613144S, PVU613151S, PVU613152S, PVU613153S, PVU613159S, PVU613170S, PVU613171S, PVU613177S, PVU613178S, PVU613187S, PVU613200S, PVU613206S, PVU613237S.