Dimension Vista System CHEM 1 CAL, Catalog Number KC110, Lot Number 9DM001 The Dimension Vista CHEM 1 CAL is...
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista System CHEM 1 CAL, Catalog Number KC110, Lot Number 9DM001 The Dimension Vista CHEM 1 CAL is an in vitro diagnostic product for the calibration... on . The reason FDA records is "Siemens Healthcare Diagnostics, Inc. has confirmed high accuracy shifts with Dimension Vista URCA Flex reagent cartridge (various lots) when calibrating with some vials from CHEM 1 calibrator lot 9DM001". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0864-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista System CHEM 1 CAL, Catalog Number KC110, Lot Number 9DM001 The Dimension Vista CHEM 1 CAL is an in vitro diagnostic product for the calibration of Calcium (CA), Cholesterol (CHOL), Creatinine (CREA), Glucose (GLU), Lactic Acid (LA), Magnesium (MG), Thyroxine (T4), Thyronine Uptake (TU), Blood Urea Nitrogen (BUN) and Uric Acid (URCA) methods on the Dimension Vista System
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- Siemens Healthcare Diagnostics, Inc. has confirmed high accuracy shifts with Dimension Vista URCA Flex reagent cartridge (various lots) when calibrating with some vials from CHEM 1 calibrator lot 9DM001.
- Root cause tag
- manufacturing process
- FDA root cause
- Other
- Action
- An Urgent Field Safety Notice dated August 31, 2009, was issued to all affected customers. Customers were instructed to discontinue the use of Dimension Vista System CHEM 1 Cal lot 9DM001 and discard any remaining product. If customers have any question they can contact Technical Solutions at 1-800-241-0420.
- Quantity in commerce
- 1474 Cartons
- Distribution
- Nationwide and Foreign: Australia, Canada, Denmark, and Malaysia.
- Product codes and lots
- Lot Number 9DM001