Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0121-2010

Siemens ARTISTE syngo RT Therapist, Therapist Assist, Therapist Express Basic, and Therapist Express Assist...

Siemens Medical Solutions USA, Inc. / initiated 2009-09-10 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Siemens ARTISTE syngo RT Therapist, Therapist Assist, Therapist Express Basic, and Therapist Express Assist Systems, manufactured by Siemens Medical Solutions... on . The reason FDA records is "A failed data transfer is possible that may result in data not being recorded in the interfaced system. Siemens Medical Solutions, Inc. USA became aware of this potential issue with its ARTISTE Linear Accelerator's...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0121-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens ARTISTE syngo RT Therapist, Therapist Assist, Therapist Express Basic, and Therapist Express Assist Systems, manufactured by Siemens Medical Solutions, Concord, CA
Product code
IYE Accelerator, Linear, Medical
Reason for recall
A failed data transfer is possible that may result in data not being recorded in the interfaced system. Siemens Medical Solutions, Inc. USA became aware of this potential issue with its ARTISTE Linear Accelerator's Syngo RT Therapist v4.1 Workspace (containing the Siemens brand of Verify and Record application called PRIMEVIEW) and the LANTIS Oncology Information System product.
Root cause tag
software
FDA root cause
Software design
Action
The firm issued a TH015/09/S Safety Modification syngo RT Therapist 4.1 Customer Advisory letter, dated September 10. 2009. the letter stated the reason for modification, the affected systems, working time, deadline, reimbursement, and reporting. The firm is preparing a software upgrade for the affected products to correct the problem. The letter was sent via certified mail.
Quantity in commerce
97 units
Distribution
Worldwide Distribution -- USA, Spain, Germany, Poland, Australia, Malaysia, New Zealand, the UK, India, Ireland, Egypt, Italy, Belgium, Republic of Korea, and Canada.
Product codes and lots
Product: syngo RT Therapist model/part #8162815 Product: syngo RT Therapist Assist model/part #8162807 Product: syngo RT Therapist Express Basic model/part #8151289 Product: syngo RT Therapist Express Assist model/part #8151289