Z-0267-2013Class II

Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different...

Medacta USA, Inc. / initiated 2008-10-13 / Terminated / root cause: manufacturing process

Medacta Usa Inc began a recall of Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in... on . FDA classified it as Class II and gives the reason as "Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0267-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medacta USA, Inc.
Product
Quadra S Offset Broach Handle The offset broach handle is intended to attach to the broaches of different sizes as the surgeon broaches the femoral bone in preparation for the implantation of the femoral stem.
Product code
JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented
Reason for recall
Medacta USA Inc has initiated a recall on the Quadra S Offset Broach Handle because of the potential of the rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Medacta USA, sent a Urgent Safety Notification letter dated October 13, 2008, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. The letter informed the customers of the lot numbers on recall and the reference numbers. The recall was initiated because the offset broach handles contain a lever mechanism that is connected to the locking axis by means of a metallic wire. The rivet that locks the metallic wire to the lever may become loose during a surgery and make it impossible to lock/unlock the broaches. For further questions please call (858) 705-0350.
Quantity in commerce
12 units
Distribution
Worldwide Distribution -- US including the states of Texas, Nevada and Colorado., and the country of Belgium.
Product codes and lots
Ref: 01.10.10.122 (Lot # 085119), Ref: 01.10.10.123 (Lot # 085120), Ref: 01.10.10.124 (Lot # 085244), Ref: 01.10.10.125 (Lot # 085243)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0267-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0267-2013Class IITerminatedVoluntary: Firm initiated