Z-0861-2010
Huber Plus Non-Coring Needle Safety Infusion Set, 19g, 1 inch, Needless Y-site, REF 011901NY, Sterile EO...
Bard Access Systems, Inc. / initiated 2010-01-14 / Terminated
Bard Access Systems began a recall of Huber Plus Non-Coring Needle Safety Infusion Set, 19g, 1 inch, Needless Y-site, REF 011901NY, Sterile EO, Assembled In Mexico, Bard Access Systems, Inc., SLC... on . The reason FDA records is "After activation of the safety mechanism, the tip of the needle protrudes from the safety sheath creating the potential for needlestick injury". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0861-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Bard Access Systems, Inc.
- Product
- Huber Plus Non-Coring Needle Safety Infusion Set, 19g, 1 inch, Needless Y-site, REF 011901NY, Sterile EO, Assembled In Mexico, Bard Access Systems, Inc., SLC, UT 84116. Safety IV administration set used to access surgically implanted vascular ports.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- After activation of the safety mechanism, the tip of the needle protrudes from the safety sheath creating the potential for needlestick injury.
- FDA root cause
- Component change control
- Action
- Bard Access Systems, Inc. issued and "Urgent: Huber Plus Non-Coring Needle Safety Infusion Set Recall" notification beginning January 2010. Users were instructed to quarantine all affected product and to notify any of their customers that had received the product. Quarantined product is to be returned to the firm. For further information, contact Bard Access Systems, Inc. at 1-800-290-1689.
- Quantity in commerce
- 4,975 units
- Distribution
- Worldwide Distribution -- United States (VA, KY, MA, NJ and TN) and Belgium.
- Product codes and lots
- Lot Numbers: RETK0229, RETJ0655.