Accolade System Surgical Technique; Accolade C - Cemented Hip Stem; Accolade TMZF Cementless Hip System...
Stryker Howmedica Osteonics Corp. began a recall of Accolade System Surgical Technique; Accolade C - Cemented Hip Stem; Accolade TMZF Cementless Hip System; Accolade HFx Cementless Hip System. Stryker... on . The reason FDA records is "Certain surgical protocols do not provide specific instructions for initiating power to the calcar planer prior to engagement with bone". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0504-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Accolade System Surgical Technique; Accolade C - Cemented Hip Stem; Accolade TMZF Cementless Hip System; Accolade HFx Cementless Hip System. Stryker Orthopaedics. Surgical protocols instruct surgeons on proper surgical technique.
- Product code
- LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented
- Reason for recall
- Certain surgical protocols do not provide specific instructions for initiating power to the calcar planer prior to engagement with bone.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- "Urgent Product Correction - Surgical Protocol" letters and Product Accountability Forms were sent to consignees via Fed Ex on August 31, 2009 and September 1, 2009 with return receipt. Notification letters described the issue, protocols affected, potential hazard and harm and risk mitigation. Direct questions about the recall by calling 1-201-972-2100 or Customer Service 1 -866-OR-ASSIST (1 -866-672-7747).
- Quantity in commerce
- 5,670
- Distribution
- Worldwide Distribution.
- Product codes and lots
- Catalog Number/Lot Code: LASST.