Z-0214-2010
Zimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, right, 9...
Zimmer GmbH / initiated 2009-10-08 / Terminated
Zimmer Inc. began a recall of Zimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, right, 9 MM height, for use with baseplate size 00 or 0... on . The reason FDA records is "The surgeon may have difficulty inserting the polyethylene into the tibial baseplate, which may result in intraoperative damage to the device". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0214-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Gender Solutions Natural-Knee Flex System, Prolong Articular Surface, congruent, Size 00,0, right, 9 MM height, for use with baseplate size 00 or 0, sterile, Zimmer, Warsaw, IN; REF 00-5424-020-09. The device is a semiconstrained total knee prosthesis components indicated for uncemented or cemented use in skeletally mature individuals.
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- The surgeon may have difficulty inserting the polyethylene into the tibial baseplate, which may result in intraoperative damage to the device.
- FDA root cause
- Other
- Action
- An "Urgent: Field Correction" dated October 8, 2009 was issued to Consignees. The notification letter described the affected product, issue, health risks and actions for customers to take. The letter notified consignees of a changed to the surgical technique, to include the use of a new peripheral rib impactor tool. Notifications of this field notification were sent to all affected direct accounts of Zimmer, Inc. Contact your Zimmer Sales Representative if you have any questions regarding this action. For shipping assistance, questions or assistance in notifying your accounts about the field correction please contact Zimmer, Inc. at 1-800-613-6131.
- Quantity in commerce
- 26,404 of all products
- Distribution
- Worldwide Distribution -- United States, Austria, Belgium, France, Germany, Japan, Spain and Switzerland.
- Product codes and lots
- Lots 60943109, 60943110, 60947725, 60954156, 60958274, 60971777, 60971783, 60988222, 60992939, 61001551, 61005435, 61012256, 61017695, 61027916, 61045079, 61051474, 61058612, 61058613, 61069414, 61075253, 61085871, 61091194, 61096357, 61106556, 61114165, 61114178 and 61136937.