Penumbra Neuron Delivery Catheter 070, percutaneous catheter, catalog numbers PND6F0701058, PND6F070956...
Penumbra Inc. began a recall of Penumbra Neuron Delivery Catheter 070, percutaneous catheter, catalog numbers PND6F0701058, PND6F070956, PND6F0701058M, and PND6F070956M; manufactured by... on . The reason FDA records is "Failure-- product can kink or ovalize, which may result in difficulty in catheter advancement or delivery of other devices and damage to catheter, which could lead to patient injury or death as a result of portions...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0001-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Penumbra Inc.
- Product
- Penumbra Neuron Delivery Catheter 070, percutaneous catheter, catalog numbers PND6F0701058, PND6F070956, PND6F0701058M, and PND6F070956M; manufactured by Penumbra, Inc., Alameda, CA. Indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- Failure-- product can kink or ovalize, which may result in difficulty in catheter advancement or delivery of other devices and damage to catheter, which could lead to patient injury or death as a result of portions being retained in the patient.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The firm issued an Urgent Voluntary Field Removal Notice letter on 09/22/2009 to its consignees, asking them to inspect their stocks; remove affected product and place in quarantine for return to Penumbra Inc.; and complete and return the attached product identification / return form.
- Quantity in commerce
- 2246 units
- Distribution
- Worldwide Distribution -- USA, Canada, Hong Kong, Chile, the UK, Denmark, Norway, Sweden, Finland, Netherlands, Germany, Switzerland, Austria, and Spain.
- Product codes and lots
- Lot numbers F13733, F13751, F13773, F13799, F13817, F13826, F13856, F13865, F13880, F13899, F13934, F13960, F13970, F13971, F13980, F13997, F14007, F14028, F14045, F14046, F14047, F14048, F14102, F14103, F14150, F14182, F14183, F14427, F14482, F14560, and F14629.