Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0525-2010

Abbott Spine Thoracolumbar Common, SpdLnk 5.5 Trans Cntr Med 43-51 mm, non-sterile, Manufactured by Abbott...

Zimmer GmbH / initiated 2009-09-21 / Terminated / root cause: manufacturing process

Zimmer Inc. began a recall of Abbott Spine Thoracolumbar Common, SpdLnk 5.5 Trans Cntr Med 43-51 mm, non-sterile, Manufactured by Abbott Spine, Inc., Austin, TX; Ref. 1302-55M. Note: This... on . The reason FDA records is "The titanium stock used to manufacturer the product is reported to exhibit areas of alloy segregation, which could result in a weakening of the strength of the metal". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0525-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Abbott Spine Thoracolumbar Common, SpdLnk 5.5 Trans Cntr Med 43-51 mm, non-sterile, Manufactured by Abbott Spine, Inc., Austin, TX; Ref. 1302-55M. Note: This lot was not distributed in the United States. The component is used in the stabilization of constructs in the BacFix and InCompass spinal fixation systems. It is used within a temporary construct that assists normal healing and is not intended to replace normal body structures.
Product code
KWP Appliance, Fixation, Spinal Interlaminal
Reason for recall
The titanium stock used to manufacturer the product is reported to exhibit areas of alloy segregation, which could result in a weakening of the strength of the metal.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Consignees were notified by letter dated September 21, 2009 and instructed to return the affected product to Zimmer, Inc.
Quantity in commerce
13
Distribution
Worldwide Distribution -- Canada, Mexico and South Africa. (Product not distributed in the United States)
Product codes and lots
Lot 55QJ.