VARiS 6.2, 6.1 and 6.0 when used with LINK, Model numbers H88 and H85. Manufactured by Varian Medical...
Varian Medical Systems Oncology Systems began a recall of VARiS 6.2, 6.1 and 6.0 when used with LINK, Model numbers H88 and H85. Manufactured by Varian Medical Systems, Palo Alto, CA on . The reason FDA records is "Software issue: Under certain circumstances, when a therapy treatment field is transferred to VARiS Vision 6.2, 6.1 and 6.0 via LINK procedures, the rotation direction will be reversed". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0133-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- VARiS 6.2, 6.1 and 6.0 when used with LINK, Model numbers H88 and H85. Manufactured by Varian Medical Systems, Palo Alto, CA.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Software issue: Under certain circumstances, when a therapy treatment field is transferred to VARiS Vision 6.2, 6.1 and 6.0 via LINK procedures, the rotation direction will be reversed.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Recall initiated September 15, 2009. An "Urgent - Field Safety Notice" letter was distributed to all affected users, with a description of the problem and corrective action steps. The letter was also distributed to Varian sales, marketing, and service organizations, informing them of the issue. For further information in the USA and Canada, contact Varian at 1-650-424-5731 or 1-888-VARIAN5 (888- 827-4265). In Europe, contact +41 41 749 8844.
- Quantity in commerce
- 100 units
- Distribution
- Worldwide distribution to 100 consignees throughout the US in IL, NY, MI, NJ, AL, TN, MD, OH, FL, and TX, and to Brazil, Chile, China, Czech Republic, Egypt, Germany, Hungary, Israel, Italy, Japan, Malaysia, the Netherlands, Romania, Slovakia, Slovenia, South Korea, Spain, Taiwan, Thailand and the United Kingdom.
- Product codes and lots
- Codes H850008, H850904, H851353, H851333, H880033, H850907, H851388, H881354, H850078, H850912, H851674, H881389, H850110, H850920, H851838, H881392, H850129, H850930, H851992, H881393, H850130, H850962, H852053, H881410, H850211, H850963, H852544, H881482, H850217, H850964, H852546, H881483, H850258, H850965, H880004, H881484, H850322, H850967, H880012, H881515, H850345, H850968, H880098, H881524, H850350, H850969, H880221, H881557, H850451, H851020, H880449, H881576, H850474, H851025, H880699, H881586, H850555, H851032, H880720, H881641, H850616, H851059, H880734, H881642, H850686, H851080, H880809, H881658, H850696, H851087, H880862, H881671, H850739, H851091, H881033, H881839, H850759, H851132, H881037, H881865, H850788, H851138, H881060, H881895, H850830, H851145, H881102, H881951, H850831, H851252, H881205, H881993, H850861, H851290, H881207, H882003, H850877, H851327, H881220, H882049, H882623, H885074