Sealapex Express (Part Number: 33639), root canal filling material. Manufactured by SybronEndo, Glendora, CA...
Sybron Dental Specialties began a recall of Sealapex Express (Part Number: 33639), root canal filling material. Manufactured by SybronEndo, Glendora, CA. Intended as a root canal filling material that is... on . The reason FDA records is "Sybron Dental Specialties is initiating a recall on the Sealapex Xpress due to the presence of crystals in the catalyst. This crystallization may affect proper canal filling by preventing Gutta Percha Points from...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0917-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Sybron Dental Specialties, Inc.
- Product
- Sealapex Express (Part Number: 33639), root canal filling material. Manufactured by SybronEndo, Glendora, CA. Intended as a root canal filling material that is used in conjunction with gutta percha or silver endodontic points.
- Product code
- KIF Resin, Root Canal Filling
- Reason for recall
- Sybron Dental Specialties is initiating a recall on the Sealapex Xpress due to the presence of crystals in the catalyst. This crystallization may affect proper canal filling by preventing Gutta Percha Points from reaching the bottom of the root canal which may lead to an inadequate seal.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- SybronEndo initiated an "Urgent Medical Device Recall" notification dated February 19 , 2009. Consignees were instructed to complete a "return form" and return any affected product in their inventory to the firm. Any product returned will be replaced at no charge. For further information, contact SybronEndo Customer Care at 1-800-346-3636.
- Quantity in commerce
- 1309 units
- Distribution
- Worldwide Distribution -- United States, Australia, Canada, United Arab Emirates, Belgium, Bulgaria, Germany, Spain, France, UK, Greece, Italy, Kuwait, Lebanon, Portugal, Russia, Sweden, Hong Kong, India, Malaysia and Taiwan.
- Product codes and lots
- Lot Numbers: 6-1314, 7-1221 and 7-1258.
FIRST ARTICLE / WEEKLY