Olsen Medical, MIDAS TOUCH NEEDLE ELECTRODE, WITH MODIFIED TIP. Model Numbers: 30-1011, 30-1012, 30-1013...
Kentucky Packaging Service LP began a recall of Olsen Medical, MIDAS TOUCH NEEDLE ELECTRODE, WITH MODIFIED TIP. Model Numbers: 30-1011, 30-1012, 30-1013, 30-1014 and 30-6011 on . The reason FDA records is "THERE WERE INCONSISTENCIES NOTICED ON THE INSULATION OVERLAP. THE ELECTRODE HAS A LARGE DIAMETER SHAFT THAT TRANSITIONS TO A SMALLER DIAMETER SHAFT. THE 1ST LAYER OF INSULATION WHICH COVERS THE SMALL DIAMETER SHAFT...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0140-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Olsen Medical
- Product
- Olsen Medical, MIDAS TOUCH NEEDLE ELECTRODE, WITH MODIFIED TIP. Model Numbers: 30-1011, 30-1012, 30-1013, 30-1014 and 30-6011.
- Product code
- GEI Electrosurgical, Cutting & Coagulation & Accessories
- Reason for recall
- THERE WERE INCONSISTENCIES NOTICED ON THE INSULATION OVERLAP. THE ELECTRODE HAS A LARGE DIAMETER SHAFT THAT TRANSITIONS TO A SMALLER DIAMETER SHAFT. THE 1ST LAYER OF INSULATION WHICH COVERS THE SMALL DIAMETER SHAFT OVERLAPS WITH THE 2ND LAYER OF INSULATION THAT COVERS THE LARGER DIAMETER SHAFT. ON SOME ELECTRODES IT APPEARS THAT THE OVERLAP IS TOO SHORT TO COVER THE TRANSITION WHICH MAKES THIS
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Olsen Medical issued an "Urgent: Device Recall" letter to Consignees beginning September 24, 2009 . Customers were instructed to inventory their stock and for all remaining product immediately discontinue use and return. Consignees were also instructed to contact their customers and notify them of the recall. For further information, contact Olsen Medical at 1-502-772-4272.
- Quantity in commerce
- 3,617 pieces
- Distribution
- Nationwide Distribution -- AZ, FL, IA, MD, MO, NM, OR, NV, TN, TX and VA.
- Product codes and lots
- Model Number: 30-1011, Lot Numbers: 011898, 012733, 012988 and 013373; Model Number: 30-1012, Lot Number: 012244; Model Number: 30-1013, all lots; Model Number: 30-1014, Lot Numbers: 011668, 011970 and 012997; and Model Number: 30-6011, Lot Numbers: 012565, 012800 and 012983.