Z-0004-2010

GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human anatomy

GE Healthcare / initiated 2009-08-12 / Terminated / root cause: software

Ge Healthcare began a recall of GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human anatomy on . The reason FDA records is "A software defect - may result in a false indication and warning on the OEC 9900 Workstation and C-arm display that the X-Ray tube Anode is over-heated. This will also cause the displayed air kerma rate and cumulative...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0004-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human anatomy.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
A software defect - may result in a false indication and warning on the OEC 9900 Workstation and C-arm display that the X-Ray tube Anode is over-heated. This will also cause the displayed air kerma rate and cumulative air kerma to be inaccurate.
Root cause tag
software
FDA root cause
Radiation Control for Health and Safety Act
Action
An Important Electronic Product Radiation Warning letter will be sent to customers identifying the issues, interim solution, and permanent solution. Questions or concerns are to be directed to the GEHC OEC service team at 1-800-874-7378.
Quantity in commerce
1411 units total
Distribution
Nationwide and Canadian Distribution
Product codes and lots
All associated serial numbers