GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human anatomy
Ge Healthcare began a recall of GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human anatomy on . The reason FDA records is "A software defect - may result in a false indication and warning on the OEC 9900 Workstation and C-arm display that the X-Ray tube Anode is over-heated. This will also cause the displayed air kerma rate and cumulative...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0004-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GE OEC 9900 Elite ESP To produce mobile fluoroscopic images of human anatomy.
- Product code
- JAA System, X-Ray, Fluoroscopic, Image-Intensified
- Reason for recall
- A software defect - may result in a false indication and warning on the OEC 9900 Workstation and C-arm display that the X-Ray tube Anode is over-heated. This will also cause the displayed air kerma rate and cumulative air kerma to be inaccurate.
- Root cause tag
- software
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- An Important Electronic Product Radiation Warning letter will be sent to customers identifying the issues, interim solution, and permanent solution. Questions or concerns are to be directed to the GEHC OEC service team at 1-800-874-7378.
- Quantity in commerce
- 1411 units total
- Distribution
- Nationwide and Canadian Distribution
- Product codes and lots
- All associated serial numbers