Z-0119-2010
Nucleus Non-Magnetic Plug for removable magnet implants, REF Z50100, STERILE EO, Cochlear Limited
Cochlear Americas / initiated 2009-09-30 / Terminated / root cause: labeling and UDI
Cochlear Americas Inc. began a recall of Nucleus Non-Magnetic Plug for removable magnet implants, REF Z50100, STERILE EO, Cochlear Limited on . The reason FDA records is "Cochlear implant device component was mis-labeled". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0119-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cochlear Americas
- Product
- Nucleus Non-Magnetic Plug for removable magnet implants, REF Z50100, STERILE EO, Cochlear Limited.
- Product code
- MCM Implant, Cochlear
- Reason for recall
- Cochlear implant device component was mis-labeled.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- On September 28, 2009, each surgeon who received a non-magnetic plug after December 5, 2008 was sent a notification letter via courier. The letter described the mislabeling issue and actions for consignees. Customers were requested to return affected product (and the quantity discarded) to the firm and review patient records to verify the implantation of the non-magnetic plug and not the mislabled product (sterile magnetic). Direct questions about the recall by calling at 1-800-523-5798.
- Quantity in commerce
- 10 units
- Distribution
- Worldwide Distribution -- United States (AL, AZ, CA, GA, IA, NC, MA, MI, MN, MO, NY, OH, OR, PA, TN, TX and WA), Brazil, and Argentina.
- Product codes and lots
- Part No. Z-50100.
FIRST ARTICLE / WEEKLY