Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0269-2010

Norian CRS Rotary Mixer Cement 10cc sterile

Synthes GmbH / initiated 2009-09-14 / Terminated / root cause: manufacturing process

Synthes USA (HQ), Inc. began a recall of Norian CRS Rotary Mixer Cement 10cc sterile on . The reason FDA records is "Non-conforming material - increased set time. Potential clinical ramifications associated with increased set time and decreased strength include extrusion of unset material and fragmentation of cured material. The risk...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0269-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Norian CRS Rotary Mixer Cement 10cc sterile
Product code
GXP Methyl Methacrylate For Cranioplasty
Reason for recall
Non-conforming material - increased set time. Potential clinical ramifications associated with increased set time and decreased strength include extrusion of unset material and fragmentation of cured material. The risk of Norian fragmentation post cure increases with a larger defect and a thicker layer of product. Customers were advised to call Synthes at 1-800-479-6329 to obtain a Return Autho
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm issued an Urgent Medical Device Recall letter dated 9/17/09 via USPS certified mail. The letter informed customers of the problem and the need to return the product. Sales Consultants that were shipped the device were notified by teleconference on 9/14/09. Consultants were instructed to go to their accounts to remove the product from inventory. Customers were advised to call Synthes at 1-800-479-6329 to obtain a Return Authorization Number. Questions are directed to Synthes at 800-620-7025 x 5453 or 610-719-5453 or to your Synthes CMF Sales Consultant.
Quantity in commerce
15 units
Distribution
The product was shipped to medical facilities in CO, KS, MA, MO, NY, SC, TX, UT, and WA.
Product codes and lots
Catalog number 614.10.01S. Lot number N996352