Z-1019-2010

RapidPoint 400 Analyzers for Blood Gas, Electrolytes and metabolites Software Versions prior to 3.5.1. The...

Siemens Healthcare Diagnostics Inc. / initiated 2009-10-30 / Terminated / root cause: design

Siemens Healthcare Diagnostics Inc began a recall of RapidPoint 400 Analyzers for Blood Gas, Electrolytes and metabolites Software Versions prior to 3.5.1. The Siemens 400 series system is intended for point of... on . The reason FDA records is "Software: Reference sensor failure may result in inaccurate pH reuslts". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1019-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
RapidPoint 400 Analyzers for Blood Gas, Electrolytes and metabolites Software Versions prior to 3.5.1. The Siemens 400 series system is intended for point of care and laboratory testing of blood gases, electrolytes and metabolites in arterial, venous and capillary whole blood samples.
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
Software: Reference sensor failure may result in inaccurate pH reuslts
Root cause tag
design
FDA root cause
Component design/selection
Action
An "Urgent Field Safety Notice" dated September 2009 was sent via overnight mail to customers and included software Version 3.5.2 with instructions for install that will detect and prevent the reporting of potentially errant test results. The customers are requested to please install the included software update immediately and return the attached Fax back form. Siemens will provide the mandatory update to all customers with the Rapidpoint 400 /405 test system. If you have any questions, please contact your local Technical Support at 508-359-3877.
Quantity in commerce
998 units
Distribution
Worldwide distribution: USA and Australia, Austria, Bahamas, Barbados, Belgium, Bosnia Herzeg, Brazil, Canada, Canary Islands, Chile, Croatia, Czech Republic, Egypt, France, Germany, Holy See (Vatican City State) Hong Kong, Ireland, Italy, Japan, Macedonia, Malaysia, M¿xico, New Zealand Not assigned P.R. China, Poland, Portugal, Puerto Rico, Refurb.Systems U.S.A Republic Korea, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, U.A.E., United Kingdom
Product codes and lots
Software Versions prior to 3.5.1