Z-0178-2010
Zimmer Trabecular Metal Shoulder, humeral stem, 12 mm stem diameter, 130 mm stem length, 48 degrees, sterile...
Zimmer GmbH / initiated 2009-09-17 / Terminated
Zimmer Inc. began a recall of Zimmer Trabecular Metal Shoulder, humeral stem, 12 mm stem diameter, 130 mm stem length, 48 degrees, sterile, Zimmer, Warsaw, IN, REF 4348-112-13. Shoulder... on . The reason FDA records is "Some of the devices may possess a taper which will not mate with the humeral head, as intended". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0178-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Trabecular Metal Shoulder, humeral stem, 12 mm stem diameter, 130 mm stem length, 48 degrees, sterile, Zimmer, Warsaw, IN, REF 4348-112-13. Shoulder replacement surgery.
- Product code
- HSD Prosthesis, Shoulder, Hemi-, Humeral, Metallic Uncemented
- Reason for recall
- Some of the devices may possess a taper which will not mate with the humeral head, as intended.
- FDA root cause
- Other
- Action
- Zimmer distributors were notified by email on 9/19/09 to locate and return the products. Letter dated 9/29/09 were sent to the Zimmer salesforce and distributors with a letter to be given to each affected hospital account to inform them of the problem. Zimmer accounts were again asked to locate and return the devices. Hospital accounts were requested to consider the available facts when monitoring potentially affected patients.
- Quantity in commerce
- 12
- Distribution
- Nationwide.
- Product codes and lots
- Lot 60947164.
FIRST ARTICLE / WEEKLY