Z-0192-2010
Revolution Blood Pump, "revOlution Centrifugal pump Catalog# 050300000***Rx ONLY***SORIN GROUP ITALIA via...
LivaNova USA, Inc. / initiated 2009-09-25 / Terminated / root cause: packaging
Sorin Group USA, Inc. began a recall of Revolution Blood Pump, "revOlution Centrifugal pump Catalog# 050300000***Rx ONLY***SORIN GROUP ITALIA via Statale 12 Nord, 86 41037 MIRANDOLA (MO) (Italy)... on . The reason FDA records is "A weakness in the plastic packaging tray could cause the tray to crack; compromising product sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0192-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- LivaNova USA, Inc.
- Product
- Revolution Blood Pump, "revOlution Centrifugal pump Catalog# 050300000***Rx ONLY***SORIN GROUP ITALIA via Statale 12 Nord, 86 41037 MIRANDOLA (MO) (Italy) Distributed in the U.S. by: Sorin Group USA, Inc. 14401 W. 65th way Arvada (CO) 80004-3599" Intended for use in cardiopulmonary bypass procedures.
- Product code
- KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type
- Reason for recall
- A weakness in the plastic packaging tray could cause the tray to crack; compromising product sterility.
- Root cause tag
- packaging
- FDA root cause
- Packaging change control
- Action
- Customers were notified by an Urgent: Field Safety Notice, dated 9/25/09, and instructed to: check their inventory for the affected product, complete the customer response form and return or visually inspect all affected product before use. For additional information contact Sorin Group Customer Service at 1-800-650-2623.
- Quantity in commerce
- 808 units
- Distribution
- Worldwide Distribution -- US, Canada, Japan, and Europe.
- Product codes and lots
- Lot numbers: 0905220163, 0905220166, 0906240182, and 0906240185.