Z-0165-2010

* (Box label) SUCTION COAGULATOR 10FR QTY 25 ***** * (Pouch label) ULTRACLEAN TM FOOT CONTROLLED SUCTION...

ConMed Corporation / initiated 2009-09-30 / Terminated / root cause: manufacturing process

ConMed Electrosurgery began a recall of * (Box label) SUCTION COAGULATOR 10FR QTY 25 ***** * (Pouch label) ULTRACLEAN TM FOOT CONTROLLED SUCTION COAGULATOR 10 FR QTY: 1 ** STERILE ***** LOT 09GHL001... on . The reason FDA records is "One production lot of Suction Coagulators received ETO sterilization instead of gamma radiation sterilization". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0165-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ConMed Corporation
Product
* (Box label) SUCTION COAGULATOR 10FR QTY 25 ***** * (Pouch label) ULTRACLEAN TM FOOT CONTROLLED SUCTION COAGULATOR 10 FR QTY: 1 ** STERILE ***** LOT 09GHL001 July 2014 MADE IN CHINA ** CONMED CORPORATION 525 FRENCH RD, UTICA, NY USA 13502"
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
One production lot of Suction Coagulators received ETO sterilization instead of gamma radiation sterilization.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Recalling firm notified all distributors by phone from 9/30 - 10/1/09. All customers were notified by letter titled "URGENT: DEVICE RECALL" dated 09/30/2009 on 10/2/2009. Customers were instructed to check their inventory for the affected product, complete the reply form and return affected product. For questions, please contact Damaris Velez or Jeff Dickinson of ConMed Electrosurgery at 800-552-0138, ext 5225 or email succoagrecall AT conmed.com.
Quantity in commerce
2700 units
Distribution
US only. CA, FL, GA, ID, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OR, PA, SC, SD, TX, VA
Product codes and lots
Lot #: 09GHL001