Z-0263-2010
Stryker Instruments 2.1 mm Cross-Cut Fissure Bur, carbide extended, sterile; REF 2296-101-621
Stryker Corp. / initiated 2009-10-05 / Terminated / root cause: supplier
Stryker Instruments Div. of Stryker Corporation began a recall of Stryker Instruments 2.1 mm Cross-Cut Fissure Bur, carbide extended, sterile; REF 2296-101-621 on . The reason FDA records is "The cross-cuts are missing on the device, which will result in poor cutting and overheating and/or smoking of the device". As of , the recall status on record is Terminated, and MDM tags the root cause as supplier.
Recall record
- Recall number
- Z-0263-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker Instruments 2.1 mm Cross-Cut Fissure Bur, carbide extended, sterile; REF 2296-101-621.
- Product code
- DZI Drill, Bone, Powered
- Reason for recall
- The cross-cuts are missing on the device, which will result in poor cutting and overheating and/or smoking of the device.
- Root cause tag
- supplier
- FDA root cause
- Vendor change control
- Action
- Consignees were notified by letter dated 10/5/09 and requested to return the devices to Stryker. The international account was notified by email.
- Quantity in commerce
- 93
- Distribution
- USA (California, Georgia, Louisiana, Ohio, Wisconsin) and Brazil.
- Product codes and lots
- Lots 08311017 and 08336027.