Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0105-2010

Tri-State Hospital Supply Corp., Centurion Healthcare Products Premie Packs, 50, sterile, Tri-State Hospital...

Centurion Medical Products, LP / initiated 2009-09-22 / Terminated / root cause: manufacturing process

Centurion Medical Products began a recall of Tri-State Hospital Supply Corp., Centurion Healthcare Products Premie Packs, 50, sterile, Tri-State Hospital Supply Corp., Howell, MI; code no. LM110... on . The reason FDA records is "The pediatric tracheal tubes used in these kits were manufactured with an internal diameter smaller than indicated in the label, which could result in an inability to remove secretions and cause partial or complete...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0105-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Centurion Medical Products, LP
Product
Tri-State Hospital Supply Corp., Centurion Healthcare Products Premie Packs, 50, sterile, Tri-State Hospital Supply Corp., Howell, MI; code no. LM110. Aspiration tray containing Portex uncuffed pediatric sized tracheal tubes.
Product code
BTR Tube, Tracheal (W/Wo Connector)
Reason for recall
The pediatric tracheal tubes used in these kits were manufactured with an internal diameter smaller than indicated in the label, which could result in an inability to remove secretions and cause partial or complete obstruction of the airway.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The firm's sales representative was notified of the recall on September 22, 2009 via email and the sole customer was notified by letter dated September 23, 2009. All product was removed from the customer account on September 28, 2009 and returned to the recalling firm.
Quantity in commerce
650
Distribution
Illinois.
Product codes and lots
Lots 901087, 903267, 907307, 910297, 901218, 908258, 2009020990 and 2009022390.