Tri-State Hospital Supply Corp., Centurion Healthcare Products, Full Term Meconium Packs, 50, sterile...
Centurion Medical Products began a recall of Tri-State Hospital Supply Corp., Centurion Healthcare Products, Full Term Meconium Packs, 50, sterile, Tri-State Hospital Supply Corp., Howell, MI; code no... on . The reason FDA records is "The pediatric tracheal tubes used in these kits were manufactured with an internal diameter smaller than indicated in the label, which could result in an inability to remove secretions and cause partial or complete...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0106-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Centurion Medical Products, LP
- Product
- Tri-State Hospital Supply Corp., Centurion Healthcare Products, Full Term Meconium Packs, 50, sterile, Tri-State Hospital Supply Corp., Howell, MI; code no. LM115.
- Product code
- BTR Tube, Tracheal (W/Wo Connector)
- Reason for recall
- The pediatric tracheal tubes used in these kits were manufactured with an internal diameter smaller than indicated in the label, which could result in an inability to remove secretions and cause partial or complete obstruction of the airway.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- The firm's sales representative was notified of the recall on September 22, 2009 via email and the sole customer was notified by letter dated September 23, 2009. All product was removed from the customer account on September 28, 2009 and returned to the recalling firm.
- Quantity in commerce
- 750
- Distribution
- Illinois.
- Product codes and lots
- Lots 01297, 07097, 09107, 12037, 06028, 2008091501, 2008120801, 2009033001 and 2009062201.