Z-0093-2010

Wallach LL50x Cryosurgical Freezer CO2 System Model: 900007

CooperSurgical, Inc. / initiated 2009-09-29 / Terminated / root cause: design

Cooper Surgical, Inc. began a recall of Wallach LL50x Cryosurgical Freezer CO2 System Model: 900007 on . The reason FDA records is "CO2 Cryosystem thermal load may exceed the exhaust gas phase, resulting in ice formation in the exhaust valve. Should this occur, the Cryo tip may cool at a less than optimal rate, resulting in ice formation in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0093-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CooperSurgical, Inc.
Product
Wallach LL50x Cryosurgical Freezer CO2 System Model: 900007.
Product code
GEH Unit, Cryosurgical, Accessories
Reason for recall
CO2 Cryosystem thermal load may exceed the exhaust gas phase, resulting in ice formation in the exhaust valve. Should this occur, the Cryo tip may cool at a less than optimal rate, resulting in ice formation in the exhaust valve.
Root cause tag
design
FDA root cause
Device Design
Action
Cooper Surgical notified customers by letter titled "Notice of Recall - Urgent" dated September 29, 2009 advising them of the problem and requesting return of the device. International accounts were notified by letter Beginning October 1, 2009. For further information, contact Cooper Surgical by fax at 1-203-799-2002 or phone at 1-203-799-2000.
Distribution
Worldwide Distribution -- United States, Canada, Australia, Belgium, England, France, Germany, Greece, Hungary, Iceland, Italy, Ireland, Latvia, Lithuania, Netherlands, Portugal, Spain, Solvenia, Sweden, Hong Kong, Republic of China, Taiwan, Angola, Cameroon, Congo, Croatia, Egypt, Ghana, Guam, Guineae Bissau, Indonesia, Kenya, Kuwait, Lebanon, Madagascar, Mali, Mauritania, Mozambique, New Zealand, Niger, Pakistan,Qatar, Russia, Sao Tome, Saudi Arabia, Senegal, Serbia, Singapore, South Africa, St Vincents & Grenadines, Oman, Thailand, Turkey, UAE and Zimbabwe.
Product codes and lots
Serial Number Prefixes of Affected Product: A, X, 10