Z-0270-2010

Syngo Imaging XS VA60 and VA70 in combination with JPEG Compression

Siemens Medical Solutions USA, Inc. / initiated 2009-09-29 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo Imaging XS VA60 and VA70 in combination with JPEG Compression on . The reason FDA records is "Distorted image. The "Distorted" image cannot be used for diagnostic findings, and, depending on the customer scenario, a repeated exam could be necessary if the image cannot be reported on any other workplace or...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0270-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo Imaging XS VA60 and VA70 in combination with JPEG Compression.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Distorted image. The "Distorted" image cannot be used for diagnostic findings, and, depending on the customer scenario, a repeated exam could be necessary if the image cannot be reported on any other workplace or recovered from another source.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued a Customer Safety Advisory notice dated 9/14/09 to all affected customers via Update Instructions IM100/09/S. The notice informs the customer of the potential issue and provides instructions to avoid its occurrence. To avoid the potential risk of the issue, the local compression function can be disabled as described in the letter. Depending on the image data, the disabling of compression will result in increased storage usage. The issue will be resolved in the release of version VA70B. For all potentially affected customers with "JPEG Compression" license an update instruction will be released which defines the necessary corrective actions in case of "Distorted" images. The update instruction will be performed on request by the customer service group. Further questions regarding this issue are directed to the local SIEMENS Uptime Service Center or +49 (9131) 84-2056.
Quantity in commerce
28 units
Distribution
The product was shipped to medical facilities in CA, FL, GA, IL, MD, MI, MO, NE, NC, NY, OH, PA, TN, TX, and WV.
Product codes and lots
Model number 7502029 - Serial numbers: 9620, 9621, 9622, 9623, 9624, 9625, 9630, 9631, 9632, 9633, 9634, 9639, 9640, 12068, 12271, 13090, 13358, 13428, 13480, 13637, 13727, 13742, 13945, 13971, 13979, 14059, 14086, and 14960.