Z-0174-2010

Syngo Imaging XS VA60 and VA70 in combination with a NAS or HSM license

Siemens Medical Solutions USA, Inc. / initiated 2009-09-30 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of Syngo Imaging XS VA60 and VA70 in combination with a NAS or HSM license on . The reason FDA records is "Incomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0174-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Syngo Imaging XS VA60 and VA70 in combination with a NAS or HSM license
Product code
LLZ System, Image Processing, Radiological
Reason for recall
Incomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip file on the NAS or HSM will contain fewer images-those which have been archived until the time of the error/interruption.
Root cause tag
software
FDA root cause
Software design
Action
The recalling firm issued a "Customer Safety Advisory Notice" dated 9/14/09 to all affected customers via Update Instructions IM101/09/S. The notice informs the customer of the potential issue and provides instructions to avoid its occurrence. Questions regarding this issue are directed to the local SIEMENS Uptime Service Center.
Quantity in commerce
2 units
Distribution
The product was shipped to medical facilities in NY and OH
Product codes and lots
Model number 7502029 - serial number 14086 and Model number 10496279 - serial number 1418