Syngo Imaging XS VA60 and VA70 in combination with a NAS or HSM license
Siemens Medical Solutions USA, Inc began a recall of Syngo Imaging XS VA60 and VA70 in combination with a NAS or HSM license on . The reason FDA records is "Incomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0174-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Syngo Imaging XS VA60 and VA70 in combination with a NAS or HSM license
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Incomplete amount of image data; specifically, The image count in the series table of the online and archive database will display the full amount of images received at the Syngo Imaging XS as archived. However, the zip file on the NAS or HSM will contain fewer images-those which have been archived until the time of the error/interruption.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The recalling firm issued a "Customer Safety Advisory Notice" dated 9/14/09 to all affected customers via Update Instructions IM101/09/S. The notice informs the customer of the potential issue and provides instructions to avoid its occurrence. Questions regarding this issue are directed to the local SIEMENS Uptime Service Center.
- Quantity in commerce
- 2 units
- Distribution
- The product was shipped to medical facilities in NY and OH
- Product codes and lots
- Model number 7502029 - serial number 14086 and Model number 10496279 - serial number 1418