Z-0084-2010
EndoClamp aortic catheter. Label reading in part: "***CARDIOVATIONS EndoClamp aortic catheter 100 cm, REF...
Edwards Lifesciences LLC / initiated 2009-09-24 / Terminated / root cause: manufacturing process
Edwards Lifesciences Research Medical, Inc. began a recall of EndoClamp aortic catheter. Label reading in part: "***CARDIOVATIONS EndoClamp aortic catheter 100 cm, REF EC1001, Sterile EO, Sterile R, Edwards Lifesciences... on . The reason FDA records is "Balloon catheters were spontaneously rupturing during use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0084-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Edwards Lifesciences LLC
- Product
- EndoClamp aortic catheter. Label reading in part: "***CARDIOVATIONS EndoClamp aortic catheter 100 cm, REF EC1001, Sterile EO, Sterile R, Edwards Lifesciences LLC Irvine, CA***". The device is a 10.5Fr wirewound, three-lumen catheter with a balloon near its tip for occluding the ascending aorta. Indicated for use in patients undergoing cardiopulmonary bypass surgery. The catheter occludes the ascending aorta, vents the aortic root and allows delivery of cardioplegia solution to arrest the heart. The pressure lumen allows monitoring of the aortic root pressure.
- Product code
- DXC Clamp, Vascular
- Reason for recall
- Balloon catheters were spontaneously rupturing during use.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Customers were notified by an Urgent - Product Recall letter, dated 9/29/09, and instructed to check their inventory, identify any unused product and return it to the company. Customers are to contact Customer Service Organization at 1-800-424-3278 to obtain a Returned Goods Authorization number, information concerning replacement product, and if they have any questions.
- Quantity in commerce
- 2,737 total
- Distribution
- Worldwide Distribution -- US, Europe, Canada, Australia, and Thailand.
- Product codes and lots
- Model EC1001 (codes are recorded in the following format: Lot #/Expiration date): 581130/Aug-09; 582944/ Aug-09; 591423/Aug-09; 591424/Nov-09; 598990/Sept-09; 598992/Nov-09; 604411/Oct-09; 605844/Dec-09; 606674/Dec-09; 606676/Jan-10; 607104/Feb-10; 608371/Feb10; 608375/Feb-10; 608377/Feb-10; 608379/Mar-10; 618753/Mar-10; 623151/Mar-09; 624601/Mar-10; 624602/Apr-10; 628961/Apr-10; 628962/Apr-10; 630344/May-10; 634975/Jun-10; 634976/Jun-10; 643171/May-10;; 644544//Jun-10; 644545/Jul-10; 644546/Jul-09; 651860/Jul-10; 651861/Aug-10; 651862/Aug-10; and 651865/Sep-10.