Z-0461-2010
Bath Safe Adjustable Shower Seat, With Back, Catalog number 727142101
Nutech Manufacturing Corp / initiated 2009-07-09 / Terminated / root cause: design
Nutech Manufacturing Corp began a recall of Bath Safe Adjustable Shower Seat, With Back, Catalog number 727142101 on . The reason FDA records is "Safety chair used in shower/bath may not meet specifications for weight load". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0461-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Nutech Manufacturing Corp
- Product
- Bath Safe Adjustable Shower Seat, With Back, Catalog number 727142101
- Product code
- IKX Aid, Transfer
- Reason for recall
- Safety chair used in shower/bath may not meet specifications for weight load.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- The subsidiary distributor, Maddak, Inc. notified customers by A "Voluntary Product Recall Notice", phone and e-mail on 07/09/2009. The own label distributor instructed customers to return their inventory of affected products received between May 1, 2009 and July 7, 2009. Distributors were instructed to immediately quarantine then return all affected inventory to Maddak and notify their accounts. Distributors were requested to instruct dealers to display a "storefront" notice and provide consumers with a customer response form. Distributors were also requested to forward a request to dealers to post a website notice if they have one. Recovered products will be replaced.
- Quantity in commerce
- 3,627 total units
- Distribution
- Worldwide Distribution -- USA and Canada.
- Product codes and lots
- Device is not identified with a lot code. Inventories received between May 1, 2009 and July 7, 2009 are being recalled.
FIRST ARTICLE / WEEKLY