Z-0191-2010

Biomet 5.5 mm peek Allthread PEEK Knotless Anchor, knotless, sterile, Biomet sports Medicine, Inc., Ontario...

Zimmer GmbH / initiated 2009-09-09 / Terminated / root cause: design

Biomet, Inc. began a recall of Biomet 5.5 mm peek Allthread PEEK Knotless Anchor, knotless, sterile, Biomet sports Medicine, Inc., Ontario, CA; REF 904842P. Intended to be implanted for... on . The reason FDA records is "These anchors may be difficult to remove, or may not detach from the inserter shaft". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0191-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet 5.5 mm peek Allthread PEEK Knotless Anchor, knotless, sterile, Biomet sports Medicine, Inc., Ontario, CA; REF 904842P. Intended to be implanted for fixation of bone fractures and for bone reconstructions.
Product code
HWC Screw, Fixation, Bone
Reason for recall
These anchors may be difficult to remove, or may not detach from the inserter shaft.
Root cause tag
design
FDA root cause
Device Design
Action
Biomet issued an "Urgent Medical Device Recall Notice" dated September 9, 2009 informing consignees of the affected product and requesting return of the device. Distributors were instructed to notify their customers copy of the recall letter and to locate and return the products from user facilities. For further information, contact Biomet at 1-800-348-9500 or 1-574-372-3983.
Quantity in commerce
894
Distribution
Worldwide Distribution -- United States and Belgium.
Product codes and lots
All lots. Lot Numbers: 006120, 006130, 006150, 310980, 457500, 457510, 457520, 528050, 549180, 597000, 597010, 689520, 689530, 689540, 843100, 843140, 933310 and 933320.