Z-0104-2010

Pointe Scientific Liquid Glucose (HEXO) Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s...

Horiba Instruments Incorporated / initiated 2009-10-19 / Terminated / root cause: manufacturing process

Pointe Scientific, Inc. began a recall of Pointe Scientific Liquid Glucose (HEXO) Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s G7517-120,G7517-120, G7517-500, and G7517-1L; 8-G7517-120... on . The reason FDA records is "There is a loss of linearity. The product fails to maintain linearity specification of 500 mg/dL". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0104-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Pointe Scientific Liquid Glucose (HEXO) Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s G7517-120,G7517-120, G7517-500, and G7517-1L; 8-G7517-120, 8-G7517-500, 3-G7517-L, 7-G7517-500 and 7-G7517-1000. (Clinical chemistry) Pointe Scientific Liquid Glucose (HEXO) is used for the quantitative determination of glucose in serum.
Product code
CFR Hexokinase, Glucose
Reason for recall
There is a loss of linearity. The product fails to maintain linearity specification of 500 mg/dL.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
The firm issued a press release, and an amended press release, on 10/30/09. Consignees were notified by an Urgent: Medical Device Recall letter dated 10/19/09, which identified the problem; and, instructed testing laboratories to inform patients' attending physicians, and determine, with their input, whether confirmation of the previous test results will be required. Customers are to examine their inventory for any of the lot numbers listed on the letter; and, for any product that has been further distributed, those customers need to be identified and notified of this product recall by including a copy of the recall letter. Customers are to complete and return the Recall Return Response form and upon receipt, Pointe Scientific will replace any disposed of reagent. The recall was expanded to include two additional lots by letter dated 11/2/09.
Quantity in commerce
400 kits plus 59 liters and 15 vials.
Distribution
Worldwide Distribution -- USA, Israel, Mexico, South Africa, and West Indies.
Product codes and lots
All lots beginning with the numbers 814301, 820502, 823901, 826801, 829401, 814301, 820502, and 831502.