Z-0272-2010
Richard Wolf Hulka Clip; for single patient use only; a sterile tubal occluding device; 2 clips per procedure...
Richard Wolf Medical Instruments Corp. / initiated 2009-10-16 / Terminated / root cause: packaging
Richard Wolf Medical Instruments Corp. began a recall of Richard Wolf Hulka Clip; for single patient use only; a sterile tubal occluding device; 2 clips per procedure tray/pouch, 10 pouches per shipping carton... on . The reason FDA records is "The sterile pouch seal may fail to remain intact during handling rendering the Hulka clip non-sterile". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-0272-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Richard Wolf Medical Instruments Corp.
- Product
- Richard Wolf Hulka Clip; for single patient use only; a sterile tubal occluding device; 2 clips per procedure tray/pouch, 10 pouches per shipping carton; Richard Wolf Medical Instruments Corp., Vernon Hills, IL 60061; item 4986.90. Indicated for use for female sterilization (permanent contraception) by occluding the fallopian tubes
- Product code
- HGB Clip, Tubal Occlusion
- Reason for recall
- The sterile pouch seal may fail to remain intact during handling rendering the Hulka clip non-sterile.
- Root cause tag
- packaging
- FDA root cause
- Packaging
- Action
- Richard Wolfe sent recall letters dated October 16, 2009 to the direct accounts via certified mail on the same date. The accounts were informed that the lot of Hulka clips did not meet their sterile seal burst test requirements, and that the sterile seal may fail to remain intact during handling, resulting in the clips becoming contaminated or non-sterile. They were instructed to inspect their inventory for the affected lot, quarantine any product found and call Richard Wolf at 800-323-9653, ext 225 or 380 to arrange for the return of the affected product for replacement. The accounts were also requested to complete and return the enclosed response form to Richard Wolf, even if they do not have any of the affected lot in stock.
- Quantity in commerce
- 150 cartons
- Distribution
- Nationwide
- Product codes and lots
- lot number 1391, expiry date 02.2010