32CM DECATHLON GOLD FUNNEL TIP STD KIT Model: DFC27SH32 15.5 Fr polyurethane twin lumen Heparin coated...
Spire Biomedical, Inc. began a recall of 32CM DECATHLON GOLD FUNNEL TIP STD KIT Model: DFC27SH32 15.5 Fr polyurethane twin lumen Heparin coated hemodialysis catheter with separated funnel tips, cuff... on . The reason FDA records is "There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0317-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Spire Biomedical, Inc.
- Product
- 32CM DECATHLON GOLD FUNNEL TIP STD KIT Model: DFC27SH32 15.5 Fr polyurethane twin lumen Heparin coated hemodialysis catheter with separated funnel tips, cuff and side holes for chronic hemodialysis and apheresis. Kit contains catheter with stylet, 18 gauge introducer needle, 12cc syringe, J/straight 0.038 inch guidewire, 12Fr dilator, 14Fr dilator, 16Fr peel-away sheath/dilator, #11 scalpel, tunneling stylet, (4) gauze sponges, wound dressing, (2) injection sealing caps, and warning label. Latex free
- Product code
- MSD Catheter, Hemodialysis, Implanted
- Reason for recall
- There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.
- Quantity in commerce
- 567 units
- Distribution
- Worldwide distribution - Nationwide, Government. Foreign: Canada, France.
- Product codes and lots
- All lot codes