Unomedical Adult Manual Pulmonary Resuscitator, Adult Medium Mask w/ Reservoir Bag. Product ID # 779725
Unomedical began a recall of Unomedical Adult Manual Pulmonary Resuscitator, Adult Medium Mask w/ Reservoir Bag. Product ID # 779725 on . The reason FDA records is "In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0070-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Unomedical A/S
- Product
- Unomedical Adult Manual Pulmonary Resuscitator, Adult Medium Mask w/ Reservoir Bag. Product ID # 779725
- Product code
- BTM Ventilator, Emergency, Manual (Resuscitator)
- Reason for recall
- In some Manual Pulmonary Resuscitator units the duckbill valve and retention ring inside the patient valve housing may become dislodged and fall into the MPR bag/body. If this occurs, the MPR may lose its ability to generate the positive pressure necessary necessary to ventilate the patient.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Unomedical issued Press on October 6, 2009. Letters were sent to customers on October 7, 2009 via UPS - return receipt. Letters explained the reason for recall and the potential hazard to patients. The letter contains a list of all affected lot numbers as well as a picture which shows that units with a clearly visible blue ring are NOT affected. Distributors are asked to examine inventory and quarantine all affected product and return the response form, even if no recalled product is in inventory. If the product was further distributed, the letter requests that a copy of the letter be sent to all customers and for returns, follow Distributor actions listed in the letter. Retailers and end users/Hospitals/EMT's are asked to perform a count of recalled product, segregate it and return the response form. All customers are asked to call 1-800-634-6003 where they will be assigned a RGA number and given instructions on how to return product. Contact information: 8:00AM - 5:30 PM Central Time Monday through Thursday and 8:00AM to 1:00 PM on Fridays, Central Time.
- Quantity in commerce
- over 200 total lots
- Distribution
- Worldwide distribution: USA, Argentina, Canada, Denmark, France, Greece, India, Italy, Japan, Mexico, Paraguay, Peru, Poland, Russia and Vietnam.
- Product codes and lots
- Product ID # 779725, lot code: 04-35