Z-0480-2010

Maquet Accessories, Connection Bracket. Catalogue # 1180.36A0. An accessory component for the MAGNUS...

Maquet Cardiovascular Us Sales, LLC / initiated 2009-10-22 / Terminated / root cause: design

Maquet Inc. began a recall of Maquet Accessories, Connection Bracket. Catalogue # 1180.36A0. An accessory component for the MAGNUS operating table system. Used to mount head positioning... on . The reason FDA records is "Product monitoring has determined that the connection bracket shows potential mechanical play at the OR table interface for some procedures. This is due to the hole in the lock washer, which is responsible for locking...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0480-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Maquet Accessories, Connection Bracket. Catalogue # 1180.36A0. An accessory component for the MAGNUS operating table system. Used to mount head positioning accessories.
Product code
FWZ Operating Room Accessories Table Tray
Reason for recall
Product monitoring has determined that the connection bracket shows potential mechanical play at the OR table interface for some procedures. This is due to the hole in the lock washer, which is responsible for locking the accessory to the OR table, being too large.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Device Correction Letters were sent by Federal Express to affected US hospitals on October 22, 2009. The letter stated the technical description, potential hazard, and the recommended precaution. The letter asked customers to immediately verify whether their Connection Bracket is affected thru the FaxBack provided. Questions or need for additional information can be directed to customers' local MAQUET representative.
Quantity in commerce
9 units in US.
Distribution
Nationwide Distribution -- USA, including states of SC and PA.
Product codes and lots
No serial or lot numbers, just a catalogue number.