Z-0451-2010

Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.5 x 58MM, spinal implant...

Howmedica Osteonics Corp / initiated 2009-10-02 / Terminated / root cause: design

Stryker Spine began a recall of Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.5 x 58MM, spinal implant component on . The reason FDA records is "Surgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0451-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.5 x 58MM, spinal implant component
Product code
KWQ Appliance, Fixation, Spinal Intervertebral Body
Reason for recall
Surgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break at the tip.
Root cause tag
design
FDA root cause
Device Design
Action
Recall notification letters were sent to all Stryker branches, Hospital Risk Management and Surgeons on October 19, 2009 by Federal Express. Questions are to be directed to Tiffani Rogers, Regulatory Compliance Manager at 201-760-8206.
Quantity in commerce
137
Distribution
Nationwide
Product codes and lots
Catalog number 48036558, all lots