Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.5 x 64MM, spinal implant...
Stryker Spine began a recall of Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.5 x 64MM, spinal implant component on . The reason FDA records is "Surgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0453-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Thor Anterior Plating System; Non Sterile Thor Standard Long Screw 6.5 x 64MM, spinal implant component
- Product code
- KWQ Appliance, Fixation, Spinal Intervertebral Body
- Reason for recall
- Surgeons may apply an accidental axial torque to and/or over-angulate the All-In One Guide while implanting the screw. This may cause the screw not to seat properly in the locking ring and may cause the Thor screwdrivers to twist or break at the tip.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Recall notification letters were sent to all Stryker branches, Hospital Risk Management and Surgeons on October 19, 2009 by Federal Express. Questions are to be directed to Tiffani Rogers, Regulatory Compliance Manager at 201-760-8206.
- Quantity in commerce
- 157
- Distribution
- Nationwide
- Product codes and lots
- Catalog number 48036564, all lots