32CM ALTA LR GOLD STD KIT Model: ALC27SH32 ALTA GOLD 15.5 Fr polyurethane twin lumen fixed tip heparin coated...
Spire Biomedical, Inc. began a recall of 32CM ALTA LR GOLD STD KIT Model: ALC27SH32 ALTA GOLD 15.5 Fr polyurethane twin lumen fixed tip heparin coated hemodialysis catheter, cuff and side holes for... on . The reason FDA records is "There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0332-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Spire Biomedical, Inc.
- Product
- 32CM ALTA LR GOLD STD KIT Model: ALC27SH32 ALTA GOLD 15.5 Fr polyurethane twin lumen fixed tip heparin coated hemodialysis catheter, cuff and side holes for chronic hemodialysis and apheresis. Kit contains catheter, 18 gauge introducer needle, 12cc syringe, J/straight 0.038 inch x 100 cm marked guidewire, 12Fr dilator, 14Fr dilator, 16Fr valved peel-away sheath/dilator, #11 scalpel, tunneling stylet, (4) gauze sponges, wound dressing, (2) injection sealing caps, and warning label. Latex free.
- Product code
- MSD Catheter, Hemodialysis, Implanted
- Reason for recall
- There is a possibility that the catheter tubing could separate from the hub/bifurcation of the device. If this occurs, bleeding is the primary clinical complication likely to result from catheter tubing/hub separation. A second potential clinical complication is air embolism, particularly for the segment of the patient population that develops negative central venous pressure during normal respira
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Spire Medical issued an "Urgent device Recall" letter dated 10/22/09 by FEDEX to end users. Users are requested to return product. For questions, contact Spire Biomedical Customer Service at 1-877-876-8635.
- Quantity in commerce
- 666 units
- Distribution
- Worldwide distribution - Nationwide, Government. Foreign: Canada, France.
- Product codes and lots
- All lot codes