Z-1133-2010

Martel Printer The Martel Printer is an accessory to the i-STAT Portable clinical Analyzer (model 300). The...

Abbott Medical / initiated 2009-11-06 / Terminated / root cause: design

Abbott Point of Care Inc. began a recall of Martel Printer The Martel Printer is an accessory to the i-STAT Portable clinical Analyzer (model 300). The i-STAT Analyzer is used by trained medical... on . The reason FDA records is "Some Martel Printers used with the i-STAT 1 system have the potential of being damaged due to the use or rechargeable battery packs that do not contain a fuse". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1133-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Martel Printer The Martel Printer is an accessory to the i-STAT Portable clinical Analyzer (model 300). The i-STAT Analyzer is used by trained medical professionals for running a variety of clinical chemistry tests and test panels contained in i-STAT test cartridges. These tests include Hematocrit, glucose, blood urea nitrogen, sodium, potassium, chloride, ionized calcium, blood gases (oxygen, carbon dioxide, and pH), Creatinine, lactate and activated clotting time.
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
Some Martel Printers used with the i-STAT 1 system have the potential of being damaged due to the use or rechargeable battery packs that do not contain a fuse.
Root cause tag
design
FDA root cause
Component design/selection
Action
An "Urgent Recall Notice" dated October 2009 was sent to customers. The notice described the product, problem and action to be taken by the customers. The customers are requested to remove the printers from use, if they have a Martel printer with the specified serial numbers, that do not have a battery door for access to the rechargeable battery pack. The customers are ask to turn in to Abbott Point of Care or a designated repair facility, all affected Martel printers for battery pack replacement. Abbott will return the Martel printers to your facility at no charge. The customers are ask to please follow the instructions on the enclosed customer reply cared. If in doubt regarding identification of the affected Martel printers, contact Abbott Point of Care Technical Support. If you have any questions regarding this information or if removing the affected Martel printers from use will negatively impact patient care, please contact Abbott Point of Care Technical Support at 800-366-8020, option 1 or your Abbott Point of Care representative.
Quantity in commerce
6432 printers (1208 have already been returned and repaired or scrapped)
Distribution
Worldwide distribution
Product codes and lots
List number 06F23-91/111300; Abbott list number: 06F23-91 Serial numbers between 20091081 and 241248976.