Z-0420-2010

The Natus Blue Light Phototherapy is also called the "neoBLUE LED Phototherapy System". The "neoBLUE mini LED...

Natus Medical Incorporated / initiated 2009-10-01 / Terminated / root cause: design

Natus Medical Incorporated began a recall of The Natus Blue Light Phototherapy is also called the "neoBLUE LED Phototherapy System". The "neoBLUE mini LED Phototherapy System" on . The reason FDA records is "The clamp connecting the neoBLUE mini phototherapy light to the accessory pole stand can sustain stress fractures or break when excessive force is used when tightening the clamp. No injuries have been reported". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0420-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Natus Medical Incorporated
Product
The Natus Blue Light Phototherapy is also called the "neoBLUE LED Phototherapy System". The "neoBLUE mini LED Phototherapy System"
Product code
LBI Unit, Neonatal Phototherapy
Reason for recall
The clamp connecting the neoBLUE mini phototherapy light to the accessory pole stand can sustain stress fractures or break when excessive force is used when tightening the clamp. No injuries have been reported.
Root cause tag
design
FDA root cause
Device Design
Action
On 10/1/09 Natus notified domestic consignees by Urgent Product Advisory Notice letter that advised them of the issue, instructed them to inspect the clamps, and included pictures of the affected clamp. Consignees are instructed to return the recalled clamps for a new replacement to be shipped by Natus. International distribution partners were notified by letter on 10/1/2009 and instructed to identify and contact their consignees. Replacements will be sent to the distribution partner for their consignees.
Quantity in commerce
2127
Distribution
Worldwide distribution: USA, Argentina, Australia, Austria, Belgium, Bermuda, Canada, Chile, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iraq, Israel, Italy, Japan, Kenya, Kuwait, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Philippines, Poland, Portugal, PR, Qatar, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, and UK
Product codes and lots
Catalogue / Part Numbers: 010101 (110V, US, new); 010101R (110V, US, refurbish); 010102 (230V, UK, new); 010102R (230V, UK, refurbish); 010103 (230V, EUR, new); 010103R (230V, EUR, refurbish); 010104 (230V, EUR, new); 010104R (230V, EUR, refurbish).  AFFECTED SERIAL NUMBERS:  60011-60013, 60015, 60017, 60021-60029, 60031-60033, 60035, 60037, 60038, 60101-60120, 60124-60128, 60131-60324, 60327-60349, 60351-60403, 60405, 60407-60415, 60417-60507, 60509-60586, 60588-60606, 60608-60611, 60613-60845, 60847-61000, 61002-61293, 61295-61340, 61346-61574, 61577-61590, 62001-62026, 62028-62032, 62035-62145, 62147-62163, 62165-62245, 62247-62252, 62254-62309, 62311-62350, 62352-62433, 62435-62481, 62485-62634, 62636-62640, and 62642-62656.