Z-0856-2010

UniCel DxI 800 and 600 Access Immunoassay Systems, UniCel DxC 880i, 860i, 660i, and 680i SYNCHRON Access...

Siemens Healthcare Diagnostics Inc. / initiated 2009-10-28 / Terminated / root cause: software

Beckman Coulter Inc. began a recall of UniCel DxI 800 and 600 Access Immunoassay Systems, UniCel DxC 880i, 860i, 660i, and 680i SYNCHRON Access Clinical Systems with Dxl System Software versions 4.2... on . The reason FDA records is "This recall was initiated because of software issue was identified in System Software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0856-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
UniCel DxI 800 and 600 Access Immunoassay Systems, UniCel DxC 880i, 860i, 660i, and 680i SYNCHRON Access Clinical Systems with Dxl System Software versions 4.2 or 4.2.1. In vitro diagnostic device used for the quantitative, semi-quantitive or qualitative determination of various analyte concentrations found in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
This recall was initiated because of software issue was identified in System Software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red warning event and stopping sample processing, the system will post a yellow caution event and continue to operate with the dispense plate in an elevated position until the instrument
Root cause tag
software
FDA root cause
Software design
Action
Beckman Coulter issued an "Urgent: Product Corrective Action" letter dated October 28, 2009. Customers were notified of the affected product and how the issue will be resolved. For further information, contact Beckman Coulter Inc. Technical Support at 1-800-854-3633.
Quantity in commerce
377 Instruments
Distribution
Worldwide Distribution -- United States and Canada.
Product codes and lots
DxI software versions: 4.2 and 4.2.1. For labeling see UniCel DxI Operator's Guide. Part Number: 387262G