Dimension Vista V-LYTE Fluids: Standard A catalog number K820
Siemens Healthcare Diagnostics Inc. / initiated 2009-10-22 / Terminated
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista V-LYTE Fluids: Standard A catalog number K820 on . The reason FDA records is "Results can elevate after sensor replacement. Sodium urine results can be falsely elevated 15 mmol/L within 24 hours after installation of a new sensor". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1037-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista V-LYTE Fluids: Standard A catalog number K820
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Results can elevate after sensor replacement. Sodium urine results can be falsely elevated 15 mmol/L within 24 hours after installation of a new sensor.
- FDA root cause
- Other
- Action
- The recalling firm issued an Urgent Field Safety Notice dated October 2009 to all affected customers. Customers were given instruction which minimize urine specimen use and report urine specimen results when QC is acceptable. Sensors are to be replaced when urine QC is out. Replacement sensors are available by contacting the firm at 800-241-0420. Please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
- Quantity in commerce
- 150 units
- Distribution
- The product was shipped to medical facilities nationwide. The product was also shipped to Australia, Canada, Belgium, France, Germany, Italy, Japan, Malaysia, New Zealand, Netherlands, Portugal, Spain, South Korea, and Switzerland.
- Product codes and lots
- Catalog number K820