Z-1039-2010

Dimension Vista V-LYTE Fluids: Diluent catalog number K835

Siemens Healthcare Diagnostics Inc. / initiated 2009-10-22 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista V-LYTE Fluids: Diluent catalog number K835 on . The reason FDA records is "Results can elevate after sensor replacement. Sodium urine results can be falsely elevated 15 mmol/L within 24 hours after installation of a new sensor". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1039-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista V-LYTE Fluids: Diluent catalog number K835
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Results can elevate after sensor replacement. Sodium urine results can be falsely elevated 15 mmol/L within 24 hours after installation of a new sensor.
FDA root cause
Other
Action
The recalling firm issued an Urgent Field Safety Notice dated October 2009 to all affected customers. Customers were given instruction which minimize urine specimen use and report urine specimen results when QC is acceptable. Sensors are to be replaced when urine QC is out. Replacement sensors are available by contacting the firm at 800-241-0420. Please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
Quantity in commerce
150 units
Distribution
The product was shipped to medical facilities nationwide. The product was also shipped to Australia, Canada, Belgium, France, Germany, Italy, Japan, Malaysia, New Zealand, Netherlands, Portugal, Spain, South Korea, and Switzerland.
Product codes and lots
Catalog number K835