Z-0602-2010
DPX-NT series, GE Lunar DEXA bone densitometers operating with enCORE software versions 12.0, 12.1...
GE Healthcare / initiated 2009-02-01 / Terminated
Ge Healthcare began a recall of DPX-NT series, GE Lunar DEXA bone densitometers operating with enCORE software versions 12.0, 12.1, 12.2,13.0, & 13.1. The systems provide an estimate of BMD on . The reason FDA records is "Failure to properly pre-indicate the technique factors to be used during a patient scan". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0602-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- DPX-NT series, GE Lunar DEXA bone densitometers operating with enCORE software versions 12.0, 12.1, 12.2,13.0, & 13.1. The systems provide an estimate of BMD.
- Product code
- KGI Densitometer, Bone
- Reason for recall
- Failure to properly pre-indicate the technique factors to be used during a patient scan.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- An Important Electronic Product Radiation Warning letter will be sent to customers. The letter will discuss the defect description and related hazards, the affected product details, safety instructions, and product correction. Customers are to contact their local GE Healthcare service representative with questions.
- Quantity in commerce
- 1,764 total
- Distribution
- Nationwide Distribution
- Product codes and lots
- software versions 12.0, 12.1, 12.2,13.0, & 13.1.