FlexShaft, product code FS14; FlexShaft II, product code FS214; FlexShaft III, product code FS312; FlexShaft...
Surgical Devices/Covidien began a recall of FlexShaft, product code FS14; FlexShaft II, product code FS214; FlexShaft III, product code FS312; FlexShaft III Sterilization Connector Guard, product code... on . The reason FDA records is "Lack of assurance that products were manufactured under GMPs. This action is being undertaken due to certain products potentially exhibiting compromised staple formation or cutting which in some instances may result in...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0397-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- FlexShaft, product code FS14; FlexShaft II, product code FS214; FlexShaft III, product code FS312; FlexShaft III Sterilization Connector Guard, product code FS3SCG; FlexShaft II Reconditioned, product code FS214RC; Manual Release for FlexShaft II & III, product code MR14.
- Product code
- GDW Staple, Implantable
- Reason for recall
- Lack of assurance that products were manufactured under GMPs. This action is being undertaken due to certain products potentially exhibiting compromised staple formation or cutting which in some instances may result in extended surgical time for the patient.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The recalling firm issued an Urgent Medical Device Recall letter dated 9/22/09 to all customers. The letter informed customers that Covidien acquired PMI, about the problem with the products, and about the need to return any products remaining inventory. Customers were asked to return the entire inventory of Power Medical Interventions product, regardless of its expiration date. This process can be initiated by contacting Power Medical Innovations customer service at 1-866-769-3763.
- Quantity in commerce
- 1100 units
- Distribution
- The products were shipped medical facilities nationwide. The product was also shipped to distributors in Canada, Japan, Germany, Ireland, Italy, Spain, Taiwan, Belgium, Sweden, Netherlands, Israel, Portugal, India, Austria, Turkey, and Greece.
- Product codes and lots
- all codes
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
