Intelligent Reload, 45 mm Blue, product code iR45B; Intelligent Reload, 45 mm Green, product code iR45G...
Surgical Devices/Covidien began a recall of Intelligent Reload, 45 mm Blue, product code iR45B; Intelligent Reload, 45 mm Green, product code iR45G; Intelligent Reload Mid Cut 45mm Blue, product code... on . The reason FDA records is "Lack of assurance that products were manufactured under GMPs. This action is being undertaken due to certain products potentially exhibiting compromised staple formation or cutting which in some instances may result in...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0405-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- Intelligent Reload, 45 mm Blue, product code iR45B; Intelligent Reload, 45 mm Green, product code iR45G; Intelligent Reload Mid Cut 45mm Blue, product code iR45MB; Intelligent Reload Mid Cut 45mm Green, product code iR45MG; Intelligent Reload, 45 mm, Vascular, product code iR45V; Intelligent Reload, 60 mm Blue, product code iR60B; Intelligent Reload, 60 mm Green, product code iR60G; Intelligent Reload Mid Cut 60mm Blue, product code iR60MB; Intelligent Reload Mid Cut 60mm Green, product code iR60MG; Intelligent Reload, 60 mm Blue No Knife, product code iR60B4.
- Product code
- GDW Staple, Implantable
- Reason for recall
- Lack of assurance that products were manufactured under GMPs. This action is being undertaken due to certain products potentially exhibiting compromised staple formation or cutting which in some instances may result in extended surgical time for the patient.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The recalling firm issued an Urgent Medical Device Recall letter dated 9/22/09 to all customers. The letter informed customers that Covidien acquired PMI, about the problem with the products, and about the need to return any products remaining inventory. Customers were asked to return the entire inventory of Power Medical Interventions product, regardless of its expiration date. This process can be initiated by contacting Power Medical Innovations customer service at 1-866-769-3763.
- Quantity in commerce
- 21,958units
- Distribution
- The products were shipped medical facilities nationwide. The product was also shipped to distributors in Canada, Japan, Germany, Ireland, Italy, Spain, Taiwan, Belgium, Sweden, Netherlands, Israel, Portugal, India, Austria, Turkey, and Greece.
- Product codes and lots
- all codes