Z-1914-2010
GammaCoat Plasma Renin Activity 125 I RIA Kit, REF: CA-1533 (100 test kit ), CA-1553 ( 500 test kit)...
Diasorin Inc. / initiated 2009-10-12 / Terminated / root cause: manufacturing process
Diasorin Inc. began a recall of GammaCoat Plasma Renin Activity 125 I RIA Kit, REF: CA-1533 (100 test kit ), CA-1553 ( 500 test kit), DiaSorin, Stillwater, Minnesota 55082-0285. Indicated for... on . The reason FDA records is "The kit control provided with the GammaCoat Plasma Renin Activity kit was labeled with the incorrect range". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1914-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diasorin Inc.
- Product
- GammaCoat Plasma Renin Activity 125 I RIA Kit, REF: CA-1533 (100 test kit ), CA-1553 ( 500 test kit), DiaSorin, Stillwater, Minnesota 55082-0285. Indicated for the quantitative determination of plasma renin activity (PRA) by the radioimmunoassay of generated angiotension I.
- Product code
- CIB Radioimmunoassay, Angiotensin I And Renin
- Reason for recall
- The kit control provided with the GammaCoat Plasma Renin Activity kit was labeled with the incorrect range.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Consignees were Emailed or faxed a "DiaSorin Customer Notification" letter dated October 21, 2009. The letter was addressed to "Dear Valued Customer". The letter described the Product name and the problem and provided the correct range for the Renin Activity Control provided with the kit. The letter also stated that replacement product for any previous failed runs. Customers are to complete the bottom portion of the letter and return it to the firm. Questions or concerns should be directed to DiaSorin Inc., Technical Services at 1-800-328-1482 or 651-439-9710.
- Quantity in commerce
- 289 kits
- Distribution
- Worldwide Distribution -- USA, including states of CA, CT, IN, LA, MI, NC, OH, TN, UT, and WA and countries of CANADA, BELGIUM, BRAZIL, INDIA, NORWAY, PORTUGAL, SPAIN, and SWEDEN.
- Product codes and lots
- Lot numbers: 123574A, 123612A, 123612C, 123620, 123620A, 123574, 123612, and 123612B.