Animas ezManager Max Diabetes Management System version 2.0.9
Animas Corp. / initiated 2009-11-02 / Terminated / root cause: software
Animas Corporation began a recall of Animas ezManager Max Diabetes Management System version 2.0.9 on . The reason FDA records is "Printed hard copy of results is not the same as the system screens- carbohydrate to insulin ratio incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0481-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Animas Corp.
- Product
- Animas ezManager Max Diabetes Management System version 2.0.9
- Product code
- MRZ Accessories, Pump, Infusion
- Reason for recall
- Printed hard copy of results is not the same as the system screens- carbohydrate to insulin ratio incorrect.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Animas issued an Urgent Medical Device Correction letter dated 11/2/09 to the pump user and another letter to the health care professional that prescribed the pump. The letters explain the issue. The patient letter is accompanied with a CD that contains the upgraded software and instructions to follow.
- Quantity in commerce
- 20,648 units
- Distribution
- Nationwide distribution, and Canada, Cayman Islands, and Denmark.
- Product codes and lots
- version 2.0.9