Z-0481-2010

Animas ezManager Max Diabetes Management System version 2.0.9

Animas Corp. / initiated 2009-11-02 / Terminated / root cause: software

Animas Corporation began a recall of Animas ezManager Max Diabetes Management System version 2.0.9 on . The reason FDA records is "Printed hard copy of results is not the same as the system screens- carbohydrate to insulin ratio incorrect". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-0481-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Animas Corp.
Product
Animas ezManager Max Diabetes Management System version 2.0.9
Product code
MRZ Accessories, Pump, Infusion
Reason for recall
Printed hard copy of results is not the same as the system screens- carbohydrate to insulin ratio incorrect.
Root cause tag
software
FDA root cause
Software design
Action
Animas issued an Urgent Medical Device Correction letter dated 11/2/09 to the pump user and another letter to the health care professional that prescribed the pump. The letters explain the issue. The patient letter is accompanied with a CD that contains the upgraded software and instructions to follow.
Quantity in commerce
20,648 units
Distribution
Nationwide distribution, and Canada, Cayman Islands, and Denmark.
Product codes and lots
version 2.0.9