PTS Panels CHOL+HDL Panel test strips for professional use with CardioChek P-A analyzer, Manufactured by...
Polymer Technology Systems, Inc. began a recall of PTS Panels CHOL+HDL Panel test strips for professional use with CardioChek P-A analyzer, Manufactured by Polymer Technology Systems, Inc., Indianapolis... on . The reason FDA records is "Extended expiration date; the product expired on February 12, 2009 but was labeled with a April 4, 2009 expiration date". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0975-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Polymer Technology Systems
- Product
- PTS Panels CHOL+HDL Panel test strips for professional use with CardioChek P-A analyzer, Manufactured by Polymer Technology Systems, Inc., Indianapolis, IN.Catalog # 1795. The product is used to measure total cholesterol, and HDL cholesterol, in whole blood.
- Product code
- CHH Enzymatic Esterase--Oxidase, Cholesterol
- Reason for recall
- Extended expiration date; the product expired on February 12, 2009 but was labeled with a April 4, 2009 expiration date.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified by letter dated Febuary 20, 2009, advised of the problem, and that the firm would replace product not consumed by the true expiration date. Direct questions to Polymer Technology Systems, Inc.
- Quantity in commerce
- 3,748 vials
- Distribution
- Worldwide Distribution.
- Product codes and lots
- Lot G751.